Calculate electron beam output corrections for extended source-to-surface distances (SSD) using
virtual source position and inverse square law. Essential for electron treatments with non-standard
setups, irregular surfaces, and bolus applications.
Calculation Method
Beam Parameters
Calibration SSD
Actual treatment distance
Virtual Source Data
Enter measured data to determine virtual source position or use typical values
dvs is fit to every measured output above (slope = Σxy/Σx²), not to a single
pair. A small RMS residual means the inverse-square model fits the data well.
Typical Virtual Source Positions
Energy (MeV)
10×10 cm² (cm)
15×15 cm² (cm)
20×20 cm² (cm)
25×25 cm² (cm)
6
95.0
92.5
90.0
88.0
9
92.0
89.5
87.0
85.0
12
88.0
85.5
83.0
81.0
16
84.0
81.5
79.0
77.0
20
80.0
77.5
75.0
73.0
Clinical Applications
Extended SSD treatments: When treating at distances greater than standard calibration SSD
Bolus applications: Account for increased distance when bolus is used
Irregular surfaces: Calculate effective SSD for sloped or curved surfaces
Total skin electron therapy: Large SSD treatments requiring accurate output corrections
Intraoperative RT: Variable SSDs during surgery
Important Considerations
Virtual source position varies with energy and field size
Measure outputs at multiple SSDs for accurate determination
Consider air gap corrections for distances > 5 cm from standard
Account for cone/cutout factors separately
Verify with measurements for non-standard configurations
Document virtual source positions for each energy/cone combination