Winston-Lutz Analysis Calculator

The Winston-Lutz test is the gold standard for verifying radiation isocenter accuracy. This test involves imaging a small radio-opaque sphere at multiple gantry, collimator, and couch angles to determine the mechanical and radiation isocenter coincidence.

Input Measurements

For each image, enter the displacement of the field (radiation) center relative to the BB in the imager plane: lateral (cross-plane, left–right) and longitudinal (in-plane, sup–inf), in mm.

Isocenter Visualization

Analysis Results

Maximum 2D Deviation (any image): -- mm
Mean 2D Deviation: -- mm
Gantry Isocenter Size (Ø): -- mm
Collimator Isocenter Size (Ø): -- mm
Couch Isocenter Size (Ø): -- mm
Radiation Isocenter Diameter (min. enclosing circle): -- mm

Understanding Winston-Lutz Analysis

Test Principle: A small metallic sphere (typically 3-5 mm) is positioned at the nominal isocenter using room lasers. Images are acquired at various gantry, collimator, and couch angles with a small radiation field.

Key Metrics:

Note: this analysis works in the 2D imager plane. A full 3D isocenter requires combining lateral offsets across gantry angles to reconstruct the BB position in room coordinates.

Treatment Type Recommended Tolerance Action Level
Conventional RT ≤ 2.0 mm > 3.0 mm
IMRT/VMAT ≤ 1.0 mm > 1.5 mm
SRS/SBRT ≤ 0.5 mm > 1.0 mm